On line monitoring system CEMS for ultra-low flue gas

Summary

Sy-9900 fixed pollution source on-line monitoring system is an ultra-low emission CEMS system launched by our company in combination with the emission standards of various industries inChinaand the national environmental protection specifications. The system is equipped with high-precision ultraviolet absorption analyzer in combination with advanced water removal and filtering and dedusting pretreatment system. It has many advantages, such as high measurement accuracy, low detection limit, direct measurement of no, no 2 is not interfered by water vapor and so on, especially suitable for gas monitoring in high humidity and low concentration environment. 

In the particle monitoring part, the high-temperature extraction laser front scattering principle technology is adopted, and the interference of wet flue gas condensate and water drop on particle monitoring is effectively overcome by heating sampling method.

 


Product features 

Heating type sampling probe rod to reduce the loss of gas to be measured in the sampling process. 

The pretreatment adopts acid condenser to inhibit the loss of SO2 in the process of sample gas condensation, and the loss rate is less than 3mg / m3. 

No and NO2 can be measured directly without converter. 

The analyzer adopts small range, high-precision AMETEK ultraviolet absorption gas analysis. 

Fast bypass, short system delay time. 

The system has few moving parts, stable and reliable operation, and extremely low failure rate.


clip_image003.png

紫外分析仪SO2 MONO2O2  


clip_image004.png

 高温抽取式粉尘仪  

  

技术参数

气体分析仪

测量原理

双光源多波长高分辨率不色散紫外吸收技术

分析组分

SO2, NO, NO2, NOx, O2(电化学)

测量范围

SO2:0-75mg/m³, NO:0-100mg/m³;NO2:0-100mg/m³

O2:0-25%

线性偏差

≤1%FS

漂移

≤1%FS/周

重复性

≤0.5%FS

响应时间

T90≤30S

环境温度

0-50 ˚C

供电电压

AC230V 50HZ, 200VA

测量输出

4-20mA, Modbus 485,

重量

18.2Kg